/
Slide set of 71 slides based on the chapter authored by Slide set of 71 slides based on the chapter authored by

Slide set of 71 slides based on the chapter authored by - PowerPoint Presentation

faustina-dinatale
faustina-dinatale . @faustina-dinatale
Follow
396 views
Uploaded On 2017-06-29

Slide set of 71 slides based on the chapter authored by - PPT Presentation

D McLean and J Shepherd of the IAEA publication ISBN 9789201310101 Diagnostic Radiology Physics A Handbook for Teachers and Students Objective To familiarize the student with Dental radiography Mobile Radiography and fluoroscopy DualEnergy XRay absorptiomet ID: 564656

ray radiology students physics radiology ray physics students teachers slide handbook review radiography dental image tube tomography mobile energy

Share:

Link:

Embed:

Download Presentation from below link

Download Presentation The PPT/PDF document "Slide set of 71 slides based on the chap..." is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.


Presentation Transcript

Slide1

Slide set of 71 slides based on the chapter authored byD. McLean and J. Shepherdof the IAEA publication (ISBN 978-92-0-131010-1):Diagnostic Radiology Physics: A Handbook for Teachers and Students

Objective: To familiarize the student with Dental radiography, Mobile Radiography and fluoroscopy, Dual-Energy X-Ray absorptiometry, Conventional tomography and tomosynthesis

Chapter 10: Special Topics

Slide set prepared

by

S. Edyvean Slide2

Review of Radiology Physics: A Handbook for Teachers and Students - 1.CHAPTER 10. SPECIAL TOPICS IN RADIOGRAPHY10.1. Introduction10.2. Dental radiography10.3. Mobile Radiography and fluoroscopy10.4. Dual-Energy X-Ray absorptiometry

10.5. Conventional tomography and tomosynthesisBibliographySlide3

Previous chapters covered 2- dimensional imagingLater chapters cover cross-sectional imaging (CT,MR, ultrasound)This chapter presents a number of special X ray imaging modalities and their associated techniques - forming a transition between projection and cross sectional imagingSpecial X-ray imaging techniquesDental radiographyMobile Radiography and fluoroscopyDual-Energy X-Ray absorptiometryConventional tomography and tomosynthesis10.1. INTRODUCTION

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide4

The tooth can be imaged in three waysIntra oral examination with the x-ray tube external and a bitewing film placed inside the mouthExtra oral examination where both the X-ray tube and detector is external to the patient to form an OPG A conebeam CT image 10.2. DENTAL RADIOGRAPHY 10.2.1. IntroductionReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide5

An intra oral examination with bite-wingIs the most common examination, and is a low cost techniquePlaces very small demands on X ray generation since the tooth is a low attenuation static objectThe image receptor is placed inside the mouth, and irradiated externally. 10.2. DENTAL RADIOGRAPHY 10.2.1. Introduction

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

X-ray tube

bite-wing film

teethSlide6

Orthopantomograph (OPG), Two dimensional images when radiographs of the entire set of teeth are requiredThe image receptor and the X ray source are external to the patientThe X ray beam is transmitted through the head - demanding significant X ray generation power and complex motion control for the X ray tube and image receptorImage receptors are film or digital detectors 10.2. DENTAL RADIOGRAPHY 10.2.1. Introduction

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide7

Cone-beam dental CTFor three dimensional information10.2. DENTAL RADIOGRAPHY 10.2.1. IntroductionReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide8

Intra oral radiographyThe intra oral X ray tube is a small robust device with a stationary target operating with a tube current of only a few mA Dental X ray tube with a stationary anode.

10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide9

Intra oral radiography (continued)The generator is typically very simple often with fixed tube voltage and tube current allowing output changes only by variations in exposure time. Major concerns with this device are for the stability of the tube head and the collimation of the beam. 10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide10

Intra oral radiography (continued)International standards require that the focus to the patient surface distance (FSD) be 200 mm. This is assured with the use of a collimating attachment that also restricts the beam to the region of the mouth being radiographed. 10.2. DENTAL RADIOGRAPHY 10.2.2. TechnologyReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

X-ray tube

bite-wing film

teeth

200 mmSlide11

Intra oral radiography (continued)The X ray equipment requires periodic QC checkingThe process of film processing requires more diligent attention. The unscreened film is removed from the light tight moisture protective wrapping and is processed either manually or with varying degrees of automation. Hand processing is probably most common and ideally requires control of temperature and processing time. 10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide12

Intra oral radiography (continued)For higher volume clinics this can be automated with film mounted on hangers that progress through the development, stop bath, fixation and rinse processes.Typically these devices have timing and temperature control but do not control chemical activity through replenishment.This is achieved in fully automatic processors, however these are typically restricted to major dental hospitals. 10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide13

Intra oral radiography (continued)The uncertainties in film processing are best controlled through sensitometry. Light sensitometers are rare in dentistry due to the small film format, However adequate results can be achieved by using a simple radiograph of a 3 step ‘wedge’ This can be easily manufactured by folding the lead foil found in the film wrap or purchased commercially Increasingly digital detectors are replacing film.10.2. DENTAL RADIOGRAPHY

10.2.2. TechnologyReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide14

Intra oral radiography (continued)Digital image capture can be achieved from an intensifying screen that is linked to a CCD camera through a tapered fibre optic coupling. The electronic signal can be transferred to an acquisition computer either through a direct cable or through ‘blue tooth’ radio frequency transmission.10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide15

OPG (Orthopantomograph)An OPG image is created by complex equipment where the X ray tube and image receptor assembly move in a horizontal plane around the head of the patient. OPG image of the teeth 10.2. DENTAL RADIOGRAPHY

10.2.2. TechnologyReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide16

OPG (Orthopantomograph) (continued)A narrow beam of radiation is formed by the tube collimationthe image receptor moves within the assembly behind a lead aperture10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

The basic movements of the OPG unit around the mandible are illustratedSlide17

OPG (Orthopantomograph) (continued)The device uses the principle of tomography and more importantly the principle of panoramic photography. This process can be illustrated through consideration of the panoramic camera used in photography. Here an acquisition aperture is used to expose an image plate that is moved behind the aperture slit to capture the image of a ‘panorama’ while the camera simultaneously slowly rotates to scan a scene. 10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide18

Cone-beam CTCT imaging has been used for some time in dentistry, including the use of custom designed units for dental applications. Their use has become more widespread recently with the advent of cone beam technology10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide19

Cone-beam CT (continued)There are many cone beam CT (CBCT) models available using a variety of acquisition schemesThey have in common a flat panel detector for acquisition, typically using either DR technology or an intensifying screen with a CCD camera (see chapter 7). 10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide20

Cone-beam CT (continued)Typically a CBCT can acquire a full field of view (FOV) that covers the whole headalthough acquisitions that are restricted to the mandible with as little as 10% of full FOV are possible. The use of these lower cost CT units opens up new potentials in some areas of dental diagnosisHowever they have significantly higher dose compared to OPG procedures10.2. DENTAL RADIOGRAPHY 10.2.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide21

Since dental examinations are amongst the most numerous, the dosimetry of these procedures is of high interest. Relevant principles and measurement techniques of dosimetry can be found inChapter 21 of this handbookand in the IAEA Technical Report Series No.45710.2. DENTAL RADIOGRAPHY 10.2.3. Dental Dosimetry

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide22

There are large variations in recorded doses between X-ray facilities.A recent study in Europe showed that for an intra oral bitewing projectionthe average incident air kerma varied from 1 to 2 mGywith a corresponding KAP measurement of 20 to 40 mGy cm2.The dose in centres that use slower film would be expected to be significantly higher. Data for OPG examinations from Europe showedKAP values ranging from 40 to 150 mGy cm2

10.2. DENTAL RADIOGRAPHY 10.2.3. Dental Dosimetry

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide23

The estimation of a population effective dose is difficult owing to the complex distribution of critical organsThere are few radiosensitive organs around the mandible with some exceptionsThe thyroid, red bone marrow, brain, and salivary glands10.2. DENTAL RADIOGRAPHY

10.2.3. Dental DosimetryReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide24

The thyroid is the main radiosensitive organ around the mandibleWell collimated X ray units should not directly irradiate this organbut it will probably receive appreciable scattered radiation Other radiosensitive organs include the red bone marrow of the mandiblethe brain The salivary glands also need to be considered as they are extensively irradiatedThey are now included as a remainder organ in the calculation of effective dose in accordance with ICRP 103.

10.2. DENTAL RADIOGRAPHY 10.2.3. Dental Dosimetry

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide25

OPG examinations estimates of effective dose giveaverage values of ~ 7 mSv using weighting factors from ICRP 60The use of ICRP 103 weighting factors has been variously estimated to increase effective dose in dentistry by 50% to 400%. Since CBCT units operate with a large FOV, their effective doses are considerably higher than for OPGwith estimates of dose varying from 60 mSv to 550mSv, for full FOVstill considerably lower than conventional head CT with effective doses of about 2 mSv.

10.2. DENTAL RADIOGRAPHY 10.2.3. Dental Dosimetry

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide26

Mobile X ray equipment ranges from small dental units to CT and MRI units carried in a large vehicle. However this chapter is restricted to simple radiographic and fluoroscopy equipment. Mobile equipment is needed when the patient cannot be brought to a fixed installation for a radiographic examination.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.1. IntroductionReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide27

Limitations of mobile equipment relate tothe availability of a suitable electrical power supply, the size and weight of the equipment and the consequent effort required to move it. The equipment design of mobile X ray equipment is varied and innovative in order to maximise the benefit given the above constraints.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.1. Introduction

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide28

Fixed angiographic X ray machines require a capacity to draw up to 100 kW with a high current multiphase supply. Assuming no loss in the high voltage transformerthe X ray output power in the secondary circuit will equal that of the primary power drawn from the electrical supply (cf Chapter 6)Therefore a domestic single phase electric supply may typically be limited to 2.4 kWWhile low power is usually not a limitation for fluoroscopic application – this is a challenge for some radiography.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide29

One solution is to charge a capacitor which is discharged across the X ray tube – the ‘capacitor discharge’ mobile. However the tube voltage will fall rapidly during the discharge of the capacitorleading to excessive surface kerma for large patient thicknesses. It is more advantageous to have an integral battery power supply which is converted to a medium to high frequency AC signal (cf chapter 5)This leads to substantial reductions in the thickness of the coils needed in the transformer design. There is also the added advantage that it can be used when there is no electrical power supply available at the examination site.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.2. Technology

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide30

The variety of possible generator designs leads to the possibility of many types of radiographic waveforms being used in the high voltage circuit for X ray generation. This leads to varying tube outputs and beam qualities for the same radiographic settings of tube voltage and tube current (cf chapter 5)Care is therefore needed when determining dosimetric factors for mobile units.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.2. TechnologyReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide31

Image quality and general quality control for mobile X ray units generally follows that used for fixed units. The use of high fluoroscopic image quality can lead to reduced procedural time, and hence reduced radiation exposure time. An important part of image quality is the setup of viewing monitors and the ambient conditions used for operation. Every effort should be made to view monitors in low ambient lighting conditions.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.3. Image quality

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide32

Mobile X ray equipment raises concerns about occupational and the public radiation exposure, as it is not operated in a purpose-built shielded environment. Assuming all X ray equipment has been checked for tube leakage, the source of radiation of occupational concern during the procedure is scatter from the input surface of the patient.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.4. Radiation protectionReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide33

It is advised that the medical physicist take field measurements of air kerma levels due to the patient scattered radiation using a patient phantom for typical radiographic and fluoroscopic procedures. As mobile radiography may take place in environments where other patients or members of the public may be in close proximity it is essential that good communication exists between the medical physicist and the staff at the location for the radiographic procedure. These staff should attend appropriate radiation safety courses that include information about radiation risk from mobile radiography. 10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.4. Radiation protection

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide34

In many cases, such as for mobile chest radiography, the use of good radiographic practice with basic radiation protection allows safe usage in most hospital environments. Simple measurements should be made to demonstrate the safety (or otherwise) of mobile X ray equipment use.10.3. MOBILE RADIOGRAPHY AND FLUOROSCOPY 10.3.4. Radiation protectionReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide35

The principle of operation involves two imagesfrom the attenuation of a low and a high X-ray energy beamUsing special imaging equipment comprising of special beam filtering and near-perfect spatial registration of the two attenuation maps 10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

Detector(s) within gantry head

x-ray source within couchSlide36

Schematic showing the components of a DXA system. The gantry configuration shows a pencil beam systempinhole source collimatorand a single detectorThese scan the patient to acquire the attneuation data10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

. (courtesy of J. Shepherd, UCSF).

Movement of source and detector to acquire attenuation dataSlide37

Other systems with slit may source collimators and segmented line detectors are called fan-beam systems, and have the advantage of higher spatial resolution and shorter scan times10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

Get permission or re-drawSlide38

The process for determining material composition can be outlined from consideration of the total attenuation of an X-ray flux passing through a subject as represented by the following formulawhere Io is the unattenuated X-ray intensity before it passes through a N materials with a thicknesses, ti , μi is the total linear attenuation, (µ/ρ)

i is the mass attenuation coefficient of the ith material, and ξi is the i

th areal density = ρiti.

10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRY

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide39

DXA can only solve for two materials simultaneously. However, three materials may be quantified: bone, lean, and fat mass, by using three fundamental assumptions, X-ray transmission through the body for the two energy windows can be accurately described by exponential attenuation processesPixels of the human body image can describe two-components i.e. either soft tissue and bone mineral, or, when bone is not present, fat and lean mass. Thus, although DXA can only solve for two compartments within individual pixels, it can describe a 3-component model for body composition.The soft tissue overlaying the bone in the image has a composition and X-ray properties that can be predicted by the composition and X-ray properties of the tissue near, but not overlaying, the bone.

10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide40

For example - simplified DXA equations will be derived for two monochromatic X-ray exposures with different energies (a high and low energy). The full solution would require integration of the attenuation across the x-ray spectrum for each energy. The attenuation equation for each exposure results in the following two equations: where the H and L superscripts represent the high and low energy X-ray beams respectively, and the “s” and “b” subscripts represent soft tissue and bone.10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRY

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide41

The solution of these equations for the areal density of bone is given by Where RS is commonly referred to as the “ratio value” for soft tissue measured for tissue surrounding but not containing the bone.10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide42

Principle of DXA is shown with 4 intensity profilesThe high energy absorption profile is multiplied by the soft tissue R-value, Rs, which accounts for differences in high and low energy absorption of soft tissue.Rs is calculated for pixels that do not contain bone10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRY

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

(drawing courtesy of J. Shepherd, UCSF).Slide43

In order to use a DXA unit to determine bone mineral density the DXA unit must be calibrated with a phantom suitable for a particular examination, for example spine, and for a particular DXA system type. Universal phantoms that can be used between different types of systems have been developed, however the calibration of DXA units is an important practical subject essential for the viability of DXA usage. 10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

Bio-Imaging Technologies, Inc

European spine

Phantom - QRM

Hologic

Examples of standard phantoms availableSlide44

The T-score and the Z-score are parameters used.The T-score is the primary diagnostic value used for osteoporosis.The T-score is inversely related to fracture risk.By international convention, the T-score is the difference between the patient’s aBMD and a young reference aBMD in units of the population standard deviation:where SD is the standard deviation of the population of young adults.

10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRY

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide45

The Z-score is used to diagnose low bone mass in young adults and children. It is the difference between the patient’s aBMD and an age- and typically ethnicity-matched reference aBMD and standard deviations:The reference values used to calculate T and Z-scores are derived from normative databases of local populations.10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRY

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide46

More information on the standards used to calculate T and Z-scores can be found in the Postitions of the International Society for Clinical Densitometry (www.iscd.org). 10.4. DUAL-ENERGY X-RAY ABSORPTIOMETRYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide47

The usefulness of sectional images, that remove the image of unwanted overlying tissues, has been well understood since the early days of X ray imaging. The formation of such images is through an analogue process known as conventional tomography.10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESISReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide48

Conventional tomography uses the principle of image blurring to remove overlying structures from a radiological image while allowing one section of body to remain in focus. 10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. PrinciplesReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide49

During image acquisition the X ray tube is in motion The image receptor moves simultaneously in the opposite directionThe projected image in the focal plane moves in same direction as the image receptor10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. Principles

movement of image receptor

movement of X-ray tube

focal plane

movement of images

objects in focal planeSlide50

The section in focus is the focal plane Regions of the body above and below the focal plane are increasingly blurred as their distance from this plane increases.10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. PrinciplesReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

fulcrum or pivot point

movement of image receptor

movement of X-ray tube

Only the red triangle remains in the same position on the image receptor

Illustration of final image

focal planeSlide51

Conventional tomography with curved focal planeConventional tomography is easily be extended for use in dental radiography by acquiring a curved focal lane The principle is to use a variable speed for the image receptor The focal plane is known as the focal trough The variable speed gives rise to a curved curved focal trough10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. Principles

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide52

Conventional tomography with curved focal planeThe principle is that if the image receptor speed is increased the focal plane moves upwards in this example (and vice versa)10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. PrinciplesReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide53

For the curved focal trough the image receptor speed changes during motionIn this exampleThe X ray tube moves at constant speed to the right The image receptor accelerates to the left during motionConsequently the focal plane moves away from the image receptor.Note the thickness of the focal trough changes in accordance to distance from the image receptor10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. Principles

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

focal plane

Accelerating movement of image receptor

X-ray tube moving at constant speed

focal troughSlide54

Tomosynthesis - a development of conventional tomography with the use of digital technology to ‘digitally’ change the speed of the image receptorIn this case one acquisition run might consist of 10 individual X ray images each read and erased in sequence throughout the one tube movement. The images are digitally added to reconstruct different focal planes in the body. It can be seen that the focal plane can be altered by advancing or retarding each image in the series by an increasing amount. 10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. Principles

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide55

Illustration of TomosynthesisThe X ray tube moves at a constant speed to the right The image receptor moves at a constant speed to the left.In this figure 4 samplings of the image are shown at tube positions A, B, C and D.Tomographic images focused on planes I, II and III are created by combining the 4 sampled images with appropriate offsets

10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. Principles

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide56

Tomosynthesis is a method for performing high-resolution limited-angle tomography It can be treated as a special case of computed tomography in which data is acquired over a limited angular range. The computed image can then be obtained using the various CT reconstruction methods.10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.1. Principles

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide57

Conventional tomography has been almost completely replaced by computed tomography in the modern radiology department. Areas where it is still used areintravenous pyelograms (IVPs) where contrast in the kidney can be conveniently placed within the focal plane to allow clear visualisation of the contrast agent. This examination is largely replaced by CT.pantomographic dental radiography(Orthopantomogram - OPG)10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS 10.5.2. Tomographic applications

Review of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1

OPG

IVPSlide58

Conventional tomography requires one tube acquisition for each focal plane image or slice. Therefore examinations requiring many slices are inherently high dose procedures. The use of tomosynthesis, on the other hand, requires only one tube motion to capture enough data to reconstruct multiple slices within the body. Today it is an emerging technology, with its most notable application so far being in mammography10.5. CONVENTIONAL TOMOGRAPHY AND TOMOSYNTHESIS

10.5.2. Tomographic applicationsReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide59

CENTRE FOR EVIDENCE-BASED PURCHASING, Digital cone beam tomography (DCBT) systems. CEP 10048; NHS PASA March 2010 [online] (2010). http://www.pasa.nhs.uk/PASAWeb/NHSprocurement/CEP/CEPproducts.htm.HELMROT, E., ALM CARLSSON, G., Measurement of radiation dose in dental radiology, Radiation Protection Dosimetry 114 (2005) 168-171.INTERNATIONAL ATOMIC ENERGY AGENCY, Guidelines for the use of DXA in measuring bone density and soft tissue body composition, Rep. TBA, IAEA, Vienna (2010).BIBLIOGRAPHYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1Slide60

LANGLAND, O.E., LANGLAIS, R.P., Principles of Dental Imaging, Williams & Wilkins, Baltimore, MA (1997).LUDLOW, J.B., DAVIES-LUDLOW, L.E., BROOKS, S.L., HOWERTON, W.B., Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT, Dentomaxillofacial Radiology 35 (2006) 219–226.WILLIAMS, J.R., MONTGOMERY, A., Measurement of dose in panoramic dental radiology, British Journal of Radiology 73 (2000) 1002-1006.BIBLIOGRAPHYReview of Radiology Physics: A Handbook for Teachers and Students - 1.1.1 Slide 1