Laks Raghupathi PhD Veerendra Shetty Siemens Corporate Development Center Bangalore India Healthcare 1 System House Algorithms Research Group Matthias Wolf PhD Siemens Medical ID: 503225
Download Presentation The PPT/PDF document "On handling low-dose CT images in the ab..." is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.
Slide1
On handling low-dose CT images in the absence of reliable DICOM header information
Laks Raghupathi Ph.D. Veerendra ShettySiemens Corporate Development Center, Bangalore, IndiaHealthcare 1 – System House, Algorithms Research Group Matthias Wolf Ph.D. Siemens Medical Solutions USA, Inc., Malvern, PA, USASYNGO R&D US Innovation
Presenter: Laks Raghupathi
March 2013 DICOM International Conference & Seminar
Copyright © Siemens AG 2013. All rights reserved.Slide2
Handling Low Dose CT
Background Capturing Dosage in DICOM Proposed Approach & Methodology Results & Discussion Author Correspondence
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT imagesin the absence of reliable DICOM header information
Copyright © Siemens AG 2013. All rights reserved.
2Slide3
Background
CT is important diagnostic imaging modality with millions of procedures performed each year Modern image acquisition technologies are intended to enable dose reduction in CT imaging while maintaining diagnostic information Evolving technology with newer scanners and reconstruction kernels. Low dose CT (LDCT) tend to have higher noise and less sharpness LDCT may affect performance of advanced post-processing algorithms, such as Computer-aided Detection (CAD)
3
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide4
Capturing Dosage in DICOM
4 The first step is to know the dosage for each imaging procedure “Exposure”, “XRayTubeCurrent”, “KV” capture input attributes CTDoseIndex field captures overall dosage Non-Type 1 attributes – not mandatory Proposal: Radiation Dose Structured Report (RDSR)
CTDIvol (mGy), DLP (mGy * cm), etc. captures dosage All new devices to encode dose in RDSR [1] Drawbacks: Old scanners may not be updated with RDSRLegacy PACS systems may not read this new SR object
__[1] D. Clunie, Extracting, Managing and Rendering DICOM Radiation Dose Information from Legacy & Contemporary CT Modalities, DICOM Intl. Conf., 2010
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide5
Proposed Approach
5 Smoothing reduces noise, but in excess removes information Algorithms like Lung CAD* expect sharp kernel Soft kernels may affect performance such as high false positive (FP) rates [2] Need to strike balance between noise and sharpness New Idea: Intelligent pre-processing filter to adjust image quality__
[2] M. Yanagawa et al., Pulmonary nodules: Effect of ASIR technique on performance of a CAD system-comparison of performance between different-dose CT scans Eur. J. Rad., 81(10) 2012
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.
__
* Lung CAD research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product.Slide6
Framework
6Scanner 1ReadingWorkplace
Intelligent Pre-Processing* [3]Post-Processing Applications (e.g., CAD)
Scanner 2
PACS
__
[3] L. Raghupathi et al., An Intelligent Pre-Processing Framework for Standardizing Medical Images for CAD and Other Post-Processing Applications,
Proc. SPIE Med.
Imag
.
, 2012
__
*Research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product.
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide7
Choosing Filters
7
Original Filter 1 Filter 2 Filter 3 Filter 4
It is possible to adjust sharpness using the filters.
Key question: Which filter is best for Lung CAD* performance?
__
* Lung CAD research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product.
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide8
Methodology
8Signal/Contrast to Noise Ratio dependent on locationUse image metrics for “adaptive” pre-processingEstimate noise (N) around air region (carina landmark)Estimate sharpness (S) in air-body borderIteratively filter till desired Noise to Sharpness ratio (NSR) reached
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide9
NSR Outputs
9OriginalFilter 1Filter 2
Filter 3Filter 4Filter 5
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide10
LDCT Details & Ground Truth
34 Siemens IRIS Kernel avg. 90 mAs, 40 solid pulmonary nodules (SPN) in 4-30mm dia. 88 Siemens IRIS Kernel simulated at 50, 100 mAs, 67 SPN in 4-30mm dia.10
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide11
Results
11Trained on large-scale CT datasetsValidation on low-dose CT images at different levelsImproved performance* (mean FP down 66%, median down 50%)
__* Lung CAD research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product.
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide12
Discussion
12 Proposed approach normalizes images with different dosage Uses measured attributes independent of acquisition parameters Demonstrated potential for improved CAD performance on LDCT Possible to extend to other post-processing applications Potential to enhance their applicability in clinical workflows Practical alternative till DICOM dosage attributes captured
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.Slide13
Acknowledgements
13Marc Rose – Healthcare1, Siemens CT DCDr. Sean Zhou and Ingo Schmuecking – SYNGO R&D US
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT imagesin the absence of reliable DICOM header informationCopyright © Siemens AG 2013.
All rights reserved.Slide14
CorrespondenceVeerendra Shetty
Veerendra.Shetty@siemens.com Head System Function House – Healthcare1Siemens Technology & Services84, Keonics Electronics City, Bangalore, India14
Thank you for your attention !
March 2013 DICOM International Conference & Seminar
L. Raghupathi et al., On handling low-dose CT images
in the absence of reliable DICOM header information
Copyright © Siemens AG 2013.
All rights reserved.