3- Part and 5- Part Analyzers DAILY Fluidic level - slide 1 of 7 3- Part and 5- Part Analyzers DAILY Fluidic level - slide 2 of 7 3- Part and 5- Part Analyzers DAILY Fluidic level - slide 3 of 7 3- Part and 5- Part Analyzers DAILY Fluidic level - slide 4 of 7 3- Part and 5- Part Analyzers DAILY Fluidic level - slide 5 of 7 3- Part and 5- Part Analyzers DAILY Fluidic level - slide 6 of 7 3- Part and 5- Part Analyzers DAILY Fluidic level - slide 7 of 7
Description: 3- Part and 5- Part Analyzers DAILY Fluidic level check then startup and shutdown Cleaning Priming (samplereagent) Valve testcheck Pressure verifications (if available) Check for condensation Weekly Waste cap replacement Cleaning of

Related Topics

Download Presentation

"3- Part and 5- Part Analyzers DAILY Fluidic level" is the property of its rightful owner. Permission is granted to download and print the materials on this website for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.

Presentation Transcript

slide1. 3- Part and 5- Part Analyzers DAILY
Fluidic level check then startup and shutdown
Cleaning & Priming (sample/reagent)
Valve test/check
Pressure verifications (if available)
Check for condensation

Weekly
Waste cap replacement
Cleaning of Sipper, Incubator, Aspiration station
Cleaning of Rinse station
Liquid flow cleaning<br>
slide2. 3- Part and 5- Part Analyzers Monthly / 2-monthly /
Pinch valve replacement
Clean all liquid containers
Clean fluidic compartments<br>
slide3. DAILY
Fluidic level check then startup and shutdown
Purge filter / bubble removal
SIP cleaning
Pressure verifications (if available)
Nozzle sonication

Monthly
Waste cap replacement FLOWCYTOMETER<br>
slide4. DAILY
Clean the apparatus with water and mild soap
Tank Clean and Rinse
Probe cleaning
Cleaning of depositions
Electrodes are properly fixed

DO NOT
Dry at high temperature
Use abrasive cleaners
Use rough clothes or brushes
Use acetone, halogenated hydrocarbons or undiluted laboratory achocols (over 30%) for cleaning Automated GEL ELECTROPHORESIS<br>
slide5. Automated GEL ELECTROPHORESIS WEEKLY
Verify the power supply voltages with Multimeter (if possible daily)
Clean the electrodes/nodes
Clean if any salt depositions found<br>
slide6. MICRO HEMATOCRIT DAILY
Use disinfectant towel or cloth with non-corrosive disinfectant for lid and other parts of centrifuge
After cleaning, dry with soft tissue or cloth

DO NOT
Use abrasive cleaners, bleach, corrosive solvents, etc…
Spill liquid inside the electronics<br>
slide7. MICRO HEMATOCRIT WEEKLY
Inspect the tube holder
Inspect the lid, lid gasket and any other wear & tear items
Apply electronic grease or white grease before re-fixing the rotor

MONTHLY
Remove the rotor and clean with disinfectant
Ensure rotor is completely dry before fixing it
Apply electronic grease or white grease before re-fixing the rotor<br>