GEM ,  Resistive M-Bulk GEM ,  Resistive M-Bulk

GEM , Resistive M-Bulk - PowerPoint Presentation

ava
ava . @ava
Follow
28 views
Uploaded On 2024-02-09

GEM , Resistive M-Bulk - PPT Presentation

ThGEM detector production Rui De Oliveira Cosimo Cantini Antonio Teixeira Olivier Pizzirusso Julien Burnens Annecy 26042012 260412 1 Rui De Oliveira CERN PCB Workshop ID: 1045874

12rui size oliveira gem size 12rui gem oliveira large mask resistive read double situationdetector situationx readoutthick gemlarge situationprotection bulk

Share:

Link:

Embed:

Download Presentation from below link

Download Presentation The PPT/PDF document "GEM , Resistive M-Bulk" is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.


Presentation Transcript

1. GEM , Resistive M-Bulk , ThGEM detector production.Rui De Oliveira, Cosimo Cantini, Antonio Teixeira, Olivier Pizzirusso, Julien BurnensAnnecy 26/04/201226/04/121Rui De Oliveira

2. CERN PCB Workshop26/04/12Rui De Oliveira218 persons Building :1000 sqr metersMaking PCBs since 1960-PCB -Rigid -Flex -Flex-rigid -Micovias -fine line (15um) -large size (up to 2m) -Thick film Hybrids -Thin film Hybrids-Chemical milling -Cu, Fe, Al, Au, Ag, W, Mb, Ti, Cr, Ni-MPGD -GEM/thinGEM/THGEM/RETHGEM -MSHP/Cobra -MICROMEGA/ Bulk/ Micro-BULK -RES BULK -Resistive MSGC -Low mass circuits -Multilayer flexes with aluminum strips -embedded heat sinks (carbon, graphite, metals, diamon)-Embedded components -passive -Active

3. GEM Large size situationDetector without spacersMicromegasLarge size situationX/Y and U/V/W readoutThick GEMLarge size situationProtection 26/04/123Rui De OliveiraOUTLINE

4. 26/04/12Rui De Oliveira4 GEM FoilPresent size 1.2m x 0.6m Future max size 2m x 0.6m Std pattern 140um pitch/70um holes

5. 26/04/12Rui De Oliveira5Base material : Polyimide 50um + 5um on both sidesPolyimide : Apical NP from company Kaneka (Japan)Supplier of the copper clad material : Nippon Mining (Japan)GEM double mask Vs GEM single MaskSame base material Hole patterning in Cu Polyimide etch Bottom electro etch Second Polyimide Etch Limited to 40cm x 40cm due toMask precision and alignmentLimited to 2m x 60cm due toBase material EquipmentDouble mask Single mask

6. 26/04/12Rui De Oliveira6GEM Double mask Vs GEM single MaskSimilar patterns , similar behavior, same material.Angles can be adjusted in both structure (Typ value : 70um copper hole , 50um polyimide hole) Steeper angles gives lower gain but also lower charging up

7. COMPASS (CERN)7GEM double mask examplesLHCb-Muon trigger (CERN)TOTEM (CERN)WindowDrift electrodeGEM stackPad Plane Support & Media-DistributionCooling Support & LV- DistributionFront-End ElectronicShielding Cover & Read-out PlaneSTAR experiment (US)26/04/12Rui De OliveiraPresent double mask production quantities : around 500 GEMs/ year in averageMax size: 40cm x 40cm

8. 8GEM Single mask examplesKLOE – Cylindrical 3 GEM DetectorGEM 800mm x 500mmRead-out 2D : 800mmx 500mmCMS RPC possible upgrade GEM 1.1m x 500mm Present production rate : 100 Gem /year (1.2mx0.6m)Expected rate for 2012 : 250 GEM/Year/technician26/04/12Rui De Oliveira

9. GEM Large size situationDetector without spacersMicromegasLarge size situationX/Y and U/V/W readoutThick GEMLarge size situationProtection 26/04/129Rui De Oliveira

10. No spacer No stretching structure: NS2 26/04/12Rui De Oliveira1010 to 15mmReadout connectorO-ring GEM attaching structure(4 pieces defining gaps)Drift electrodeFree to slide External screws to adjuststretching GNDGNDgluing

11. 26/04/12Rui De Oliveira11FRAMEGEMHV DIVIDERDRIFT BOARDCLOSEDDETECTOR30 x 30 cmScrewsO-ringHV BNC ConnectorGas connectorsRead-out connectors1m x 0.5m under production

12. No spacers in active areaAssembly time ½ hour for 10cm x 10cm detector (1 technician)2 hours for 1m x 0.6m detector (1 technician)No gluing , no solderingRe opening possibleGEM exchange possibleFull detector Re-cleaning possible No intermediate testNew GEM stretching method (NS2) advantages:26/04/1212Rui De Oliveira

13. GEM Large size situationDetector without spacersMicromegasLarge size situationX/Y and U/V/W readoutThick GEMLarge size situationProtection 26/04/1213Rui De Oliveira

14. 26/04/1214PCBlaminationMesh depositlaminationdevelopmentRui De OliveiraLarge MicromegasLargest size produced:1.5m x 0.6mLimited by equipment BULK TechnologyDUPONT PC 1025 coverlayBOPP MeshesSERITEC stretching

15. 15Fine segmentation 1cm², thickness 8mm for ILC Hadronic calorimetryTested in the RD51 1 kHz beamBulk Micromegas ILC DHCAL first m2LAPP Annecy26/04/1215Rui De OliveiraPad read-out large TPCNeutrino detector10 sqr meter in serviceT2K experimentJapan

16. Atlas CSC replacement project 26/04/12Rui De Oliveira16MPGD Vs Wire chambers-Faster-no need for fancy gases-lower cost-robust1.2m x 0.6m

17. Resistive Bulk MicroMegas structure 26/04/12Rui De Oliveira17

18. 26/04/12Rui De Oliveira18Resistive strip processPCBPCBPCBPCBPCB-Good resistor uniformity-Easy to clean-Thermally robustRead-out strips or padsCopper lamination layerResistive inverse patterningResistive layer depositionCopper stripping

19. 26/04/12Rui De Oliveira19Atlas CSC replacement project Double sided BoardResistive strip depositBulkingTest before closingClosing

20. 26/04/12Rui De Oliveira20Situation and on going productions Situation on large STD BULKUp to 5 detectors (1.5m x 0.6m max) produced all goodSituation on large resistive detector production4 CSC type (1.2m x 0.6m) 1D produced 1 goodPhysical parameters OK , leakage current out of tolerance4 MSW type (1.2m x 0.6m) 2D produced 1 goodPhysical parameters OK , leakage current out of toleranceYield 25% due to dust contamination (solution found recently)Ionic contamination and pillar shape are not the cause of leakage currentsNext production with corrections1 detector 1 m x 1m end of June

21. GEM Large size situationDetector without spacersCompact HV dividerMicromegasLarge size situationX/Y and U/V/W readoutThick GEMLarge size situationProtection 26/04/1221Rui De Oliveira

22. Resistive stripsBulk MicroMegas resistive anode and 2D readout26/04/12Rui De OliveiraWe have produced : X/Y read-out U/V/W read-out Grounded mesh detectors22

23. 26/04/12Rui De Oliveira23X/Y read-outU/V/W read-out

24. GEM Large size situationDetector without spacersCompact HV dividerMicromegasLarge size situationX/Y and U/V/W readoutThick GEMLarge size situationProtection 26/04/1224Rui De Oliveira

25. 26/04/12Rui De Oliveira25THGEM description: -800mm x 400mm Halogen Free -1mm & 2mm thick -RIM 40um (differential etching) -0.8mm pitch -0.5mm hole diameter@ ELTOS:-Drilling-Photolithography-Differential etching@ CERN: -Thermal treatment -Passivation -Cleaning -Electrical TestTHGEM and read-out for Double phase pure argon LEM-TPC with 2D anode

26. GEM Large size situationDetector without spacersCompact HV dividerMicromegasLarge size situationX/Y and U/V/W readoutThick GEMLarge size situationProtection 26/04/1226Rui De Oliveira

27. THGEM: limitations Strong ionizing particles produce sparks. Sparks damage the electrodes and are responsible of a low degradation of the gain.The active element then drains steadily a DC current from the power supply.27 Ideas to cancel the phenomenum :-increase the number of TH GEM (we want to avoid)- Reduce the capacity by segmenting the THGEM - Decouple sectors with resistors- Resistive electrodes- Change metal- Rim , no rimChange dielectricsAfter long sparking period they all failed but some structures are 1000 time better than othersWe went back to theory to understand the problem26/04/12Rui De Oliveira

28. Voltage-Current Characteristic of the DC Electrical Discharges- Gas Discharge Physics, Yuri P. Raizer- DC pulse-powered micro-discharges on planar electrodes, Yogesh B. GianchandaniBreakdowntriggeredSelf maintained28C :negligibleavalancheCoronaGlow dischargeSpark26/04/12Rui De Oliveira

29. 29V: supply voltage (2KV)R: limiting resistor (1Mohms)C: THGEM capacity (1nF)R’: spark local surface resistor (10 Ohms) : sparkI : charge currentI’: discharge current inside capacitorCharge  T = RC (1us)I max =V/R= 2mA (limited to 10uA)Spark discharge T = R’C (10ns) I’ max = (200 A)That’s consistent with measurementsCVcVII’26/04/12Rui De Oliveira

30. 30Initial Working pointR’ low valueRepetitive sparksEnd working pointR’ value > RDC currentR’ increaseDue to spark damage ex:1 MOhmsR’ is low Clean metal ex: 10 Ohms26/04/12Rui De Oliveira

31. ?26/04/12Rui De Oliveira31High energy Sparks Glow dischargePerfect Hole ?Low energy sparks?

32. Thank youQuestions?26/04/12Rui De Oliveira32