Francesca Stocker 16 January 2019 What it should be and what it is for all APAs except APA 3 APA 3 has a disconnected GPlane After putting HV back last week 09January we did not see any signals on U V X Plane of APA 3 ID: 926498

apa plane disconnected current plane apa current disconnected apas charge time compare naturally data pulse signals voltage bias 2018

Share:

Link:

Embed:

Download Presentation from below link

Download Presentation The PPT/PDF document "APA 3, disconnected g-Plane Update" is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.


Presentation Transcript

Slide1

APA 3, disconnected g-Plane Update

Francesca Stocker

16. January 2019

Slide2

What it should be and what it is for all APAs except APA 3

Slide3

APA 3 has a disconnected G-Plane

After putting HV back last week (09.January) we did not see any signals on U, V, X – Plane of APA 3.

 no transparency , something wrong with the G - Plane

Measured capacitance of G-Plane connection on APA 3 = 0.97nF

For other APAs capacitance is around 3nF

Not a power supply problem, no cable issue outside the cryostat

--> G – Plane of APA 3 is NOT connected and can not be biased with a PS

Slide4

Good News - we still see signals

We have a self healing system:

When HV is back to nominal the floating G-Plane starts collecting electrons on the wire and after some time it will become transparent and let the electrons from tracks pass through and induce signals on U, V, X.

Natural

self biasing if you want

We cannot measure the potential on the G-Plane, but it must be < -320 V (potential of U-Plane)

Slide5

Important questions to be investigated ASAP

Since WHEN is the G-Plane of APA 3 disconnected?

Only after the Christmas break?

During the whole run and we never noticed because it would charge up naturally before we looked at the data?

What is the time that it takes for the APA 3 G-Plane to charge naturally?

How do pulse shapes and

dQ

/dx of APA 3 compare to other APAs?

Slide6

2018-09-22: G-plane bias voltage @ -665V

APA 1 G plane current

APA 2 G plane current

APA 3 G plane current

1) Since WHEN is the G-Plane of APA 3 disconnected?

Slide7

2019-01-13: Last Sunday

APA 1 G plane current

APA 2 G plane current

APA 3 G plane current

Moment bias voltage was turned off

1) Since WHEN is the G-Plane of APA 3 disconnected?

Slide8

1) Since WHEN is the G-Plane of APA 3 disconnected?

Look at the pulse shape and

dQ

/dx of runs during the whole data taking, from September 2018 until now.

How do APA 3 and APA 2 (adjacent APAs) compare?

APA switch

Slide9

2) What is the time that it takes for the APA 3 G-Plane to charge naturally?

Back of the envelope estimate gives the order of 6h for the G-plane to charge up

Run 6301 (after X-mas break, just after HV back on) should contain the charging up process of the G-Plane, can someone investigate??

Find other runs before Christmas that had a longer time HV off (around October 26/27 3 days off)

Slide10

3) How do pulse shapes and dQ/dx of APA 3 compare to other APAs?

Look at data