Variation in Electrode Impedances After Cochlear Implant Surgery Variation in Electrode Impedances After Cochlear Implant Surgery

Variation in Electrode Impedances After Cochlear Implant Surgery - PowerPoint Presentation

isabella2
isabella2 . @isabella2
Follow
27 views
Uploaded On 2024-02-02

Variation in Electrode Impedances After Cochlear Implant Surgery - PPT Presentation

Yassin Abdelsamad PhD Research Dept Manager MEDEL Saudi Arabia The Team Workgroup Dr Yassin Abdelsamad Dr Salman Alhabib Dr Medhat Yousef Dr Farid Alzhrani Prof Abdulrahman Hagr ID: 1044107

surgery electrode activation impedance electrode surgery impedance activation values normal device measured variation year session yassin med performed stimulation

Share:

Link:

Embed:

Download Presentation from below link

Download Presentation The PPT/PDF document "Variation in Electrode Impedances After ..." is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.


Presentation Transcript

1. Variation in Electrode Impedances After Cochlear Implant Surgery Yassin Abdelsamad, PhDResearch Dept. Manager, MED-EL, Saudi Arabia

2. The Team / WorkgroupDr. Yassin AbdelsamadDr. Salman AlhabibDr. Medhat YousefDr. Farid AlzhraniProf. Abdulrahman Hagr

3. BackgroundCochlear Implantation (CI) is a popular surgery performed to treat the hearing impaired with profound sensorineural hearing loss.Electrode impedance is the resistance to the current flow in the perilymph when a voltage is appliedIt’s measured Intra- and Post CI surgery to confirm the device integrity and functionality. It also help in determining the appropriate levels of electric stimulation and monitoring any unexpected issues. External DeviceInternal Device

4. Goals of the project and final users that will benefitPrimary Objective: To monitor and assess the variation in the electrical impedance of CI electrode arrays at different time points up to one year after the surgery. Secondary Objective: To predict the normal variation trending so that the programmers can detect any irregular variations.Patients, Clinical Audiologists, Clinical Engineers, and CI professionals could get benefits from this work.

5. Methods and MaterialsA retrospective study in KAESC, Riyadh, Saudi Arabia 30 ears implanted with CI and received their external devices after one month from the surgery. Normal inner ear anatomy;Implanted with the same electrode array (MED-EL; Flex28); Implantation was performed through same surgical approach; Normal intraoperative measurements; Normal electrode position inside the cochlea.The electrode impedances along the 12 electrodes of electrode arrays were measured during the activation session and at 3, 6, and 12 months after the surgery.

6. ResultsIn the included subjects, the mean electrode impedance measured at the activation session was 7.15 KΩ. Then, this value keeps decreasing till the period of one year. The recorded values were 6.29, 5.98, and 5.67 KΩ at three, six, and twelve-month after device activation.

7. ResultsFor the studied cases, the mean impedance values after one year were significantly lower than the values recorded during the first activation session (20.5% lower, p < 0.05).

8. ConclusionesThe electrode impedance is very important factor to be measured during and after the CI surgery.The electrode impedance of the cochlear implant electrode array keeps decreasing gradually after the device activation.It reaches the stable values after 6 to 12 months regardless the continues increase in the electrical stimulation levels.

9. Yassin AbdelsamadYassin.samad@gmail.comCairo, Egypt