Download Presentation The PPT/PDF document "Instrument Translation Tables Urban Myt..." is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.
Instrument Translation Tables Urban Myths Debunked Peggy Steele Lab Implementation Specialist 727-789-0100 x4583 peggy@softcomputer.com
Instrument Interfaces in a Nutshell
Instrument Interfaces in a Nutshell Translation Table
4.0 4.5 Finding the Translation Table
Interface Setup
4.5 4.0
4.0 4.5
Translation Table (4.0)
Translation Table (4.5)
Translation Table (4.0) Translates Test Codes SoftLab -> Instrument Instrument -> SoftLab Interprets Which test code to use in which scenarioi.e.- Plasma vs. Serum, RED vs. LAV Defines Partial PostingDefines which test is used for QCBuilds Loadlist Defines a reflex testModifiesModify results before posting to Lab DB Defines what should Qualify
Translation Table (4.5) Translates Test Codes SoftLab -> Instrument Instrument -> SoftLab InterpretsWhich test code to use in which scenario i.e.- Plasma vs. Serum, RED vs. LAV Defines Partial PostingDefines which test is used for QCBuilds Loadlist Defines a reflex testModifies (now done in Result Adjustment Rules Tab) Modify results before posting to Lab DB Defines what should Qualify
Translates Test Codes (4.0) Translates tests ordered in the SoftLab ® into codes understood by the instrument and converts instrument test codes with results into codes understood by SoftLab® .
Translates Test Codes (4.5)
Interprets (4.0) Which test code to use in which scenario i.e.- Plasma vs. Serum, RED vs. LAV
Interprets (4.0) Which test code to use in which scenario i.e.- Plasma vs. Serum, RED vs. LAV
Interprets (4.0) Which test code to use in which scenario i.e.- Plasma vs. Serum, RED vs. LAV
Interprets (4.5) Which test code to use in which scenario i.e.- Plasma vs. Serum, RED vs. LAV
Defines Partial Posting
Defines Partial Posting
Defines which test is used for QC (4.0) Y
Defines which test is used for QC (4.5) Y
Builds Loadlist Defines what should Qualify
Defines a Reflex Test
Translation Table Review Translates Test Codes SoftLab -> Instrument Instrument -> SoftLab InterpretsWhich test code to use in which scenario i.e.- Plasma vs. Serum, RED vs. LAV Defines Partial PostingDefines which test is used for QCBuilds Loadlist Defines a reflex test
Translation Table in 4.0 Translates Test Codes SoftLab -> Instrument Instrument -> SoftLab Interprets Which test code to use in which scenarioi.e.- Plasma vs. Serum, RED vs. LAV Defines Partial PostingDefines which test is used for QCBuilds Loadlist Defines a reflex testModifies Modify results before posting to Lab DB Defines what should Qualify
Result Manipulation Options Patient DB Resulting Worklist Calculations Post Instrument Result DB Pipe and Result Adjustment Rules Instrument Menu Interface
Write Pipe Rules The usual translation has to already be on the translation table.Start a Pipe Rule with a vertical bar or pipe (|) in the Instrument ColumnLeave System, Ext and Flt blank Use the general syntax < expression><space><new value> and the specific syntax on the following page in the Description Column.
Pipe Rules (4.0) Item Description Example Rule <expression><space><new value> <10 Negative Expression <operator> <value> <10 Operators > < >= <= == (numeric comparison) = (string comparison) >10 <10 >=10 <=10 ==10 =+++ “And” <operator> <value> & <operator><value> <10&>5 “Or” <operator><value> | <operator><value> <10|>50
Result Modification <expression><space><new value>
Numeric Result Equality Interpretation
Alphanumeric Result Equality Interpretation
Change Result from Numeric to Alphanumeric
Automatic Cancellation of Result
Unconditional Change of Result
Practice Instrument System Ext Flt Description 1 2 3 4 Instrument Sends 305 for test CHOL on tube ext. 33 When Instrument sends “<=50” lab wants to see “NEGATIVE ”. Write the Pipe Rule…
Generate Errors for the Error Table
Practice Instrument Sends 305 for test CHOL When Instrument sends “<=50” lab wants to see “NEGATIVE” . Here’s the Pipe Rule… Instrument System Ext Flt Description 1 305 CHOL 33 Cholesterol 2 |305 (WRONG) <=50 NEGATIVE 3 |305 =<=50 NEGATIVE 4
Exercise The Batch Assay Instrument that you are installing will be running Rubella and Herpes IgG screens. The specimen for all the tests is a red top tube with an extension of 32. 1. The LIS code for the Rubella IgG screen is RUBIG. The Instrument code for the test is RIG. The LIS code for the Herpes IgG screen is HERIG. The Instrument code for the test is HIG . Define these in the translation table. 2. The instrument will send numeric values for the Herpes IgG screen that need to be translated to alphanumeric value. Define the interpretations as follows: • >1.200 should be interpreted as REACTIVE • >=0.800 or <=1.2 should be interpreted as EQUIVOCAL • <0.800 should be interpreted as NONREACTIVE 3. Based on the information given, fill in the translation table
Exercise Instrument System Ext Description RIGRUBIG32 Rubella IgGHIG HERIG 32 Herpes IgG |HIG >1.200 REACTIVE|HIG >=.800&<=1.200 EQUIVOCAL |HIG <.800 NONREACTIVE
Result Adjustment Rules Modifies Result before PostingInstrument Result must meet ConditionOnce Condition is met Action will occur
Conditions Condition Definition Example = Equals =POSITIVE > , < Greater Than, Less Than >5 >= , <= Greater Than or Equal To, Less Than or Equal To >=5 & “ AND ” that joins expressions together for ranges. >5&<10 Error= Error sent Error=H
Actions Command Definition Example Convert= Convert result from old value to new value. Convert=POS Error= Sets an error value Error=H Flag Sets a flag value. Currently the only available value is CANCEL. Enter CANCEL in uppercase text. Flag=CANCEL Precision= Sets precision, or number of decimal places, inside a result value. Precision=2 Factor= Multiplies a result value Factor=100 Prepend= Adds a prefix to a value Prepend => Format= Used only to change format from standard notation to scientific notation. Enter SCIENTIFIC in uppercase text Format=SCIENTIFIC ; Semicolon used to join multiple commands in one Action entry Factor=2;Precision=2
Adjustment Rules Practice
Adjustment Rules Practice What’s wrong with these rules?
TT Last Thoughts Export- in both versions, you can right click on the grid and export to Excel Tt buttons
TT Last Thoughts TT buttons
Add
Delete
Search
Check Checks the translation table for any apparent errors in the table. The main checks performed are for entries that do not include instrument codes, incorrect syntax, and incorrect extensions.
Build (4.5) Automatically fills Translation table with Reference Lab reportable codes defined in Test Setup. Only active in edit mode if the workstation belongs to a Reference Lab location and flag 62, "Use test setup instead of translation table," is checked. Warning message will display alerting the user all current settings will be overwritten by Test Setup. Confirmation message will display once all updates to the table have been completed