Download The PPT/PDF document "Studies on TH2 chamber Michele, Maria," is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.
Slide1
Studies on TH2 chamber
Michele, Maria,
Givi, JoergWeekly MicroMegas Meeting 12/3/2013
1
MicroMegas Weekly Meeting 2013/03/12
Slide2Overview
TH2 chamber Vs
Txx chambersTH2 studies with X-rayCosmic stand for 10x10 cm chambersTH2-Txx comparison with cosmic
HV scan on TH2 chamber with FR4 frame
HV scan on T3 chamber with
Alu
frame
2
MicroMegas Weekly Meeting 2013/03/12
Slide3TH2 chamber Vs
Txxx
chamber
TH2 Chamber
T3 Chamber
RO Strips with screen printing deposition
technique
~
60 µm
~
15/20 µm
3
MicroMegas Weekly Meeting 2013/03/12
Slide4Setup X-ray Gun
Electron current of the gun: ~1 mA
Monitor detector current from HV power supplyChambers tested: TH2, T3, T8, Z1
In order to evaluate the TH2 chamber gain
w.r.t
. the others, a HV scan under X-ray irradiation has been performed
4
MicroMegas Weekly Meeting 2013/03/12
Slide5Current Measurement (X-Ray)
From the HV scan with X-ray gun, it is evident that the TH2 chamber has about 10 Volts of shift w.r.t
. the T3-T8 chambers.A few microns difference in
height of the
amplification
region is enough
to explain this
difference
5
MicroMegas Weekly Meeting 2013/03/12
Slide66
MicroMegas Weekly Meeting 2013/03/12
Setup – Mini Cosmic Stand
Scintillator
1
Scintillator
2
4 cm Lead
T7 Chamber
T3 Chamber
TH2 Chamber
Slide77
MicroMegas
Weekly Meeting 2013/03/12Setup – Mini Cosmic Stand
HV scan to evaluate Cluster charge response of TH2
w.r.t
. T3/T7 chambers
HV scan replacing Aluminium frame with plastic frame on TH2 chamber
HV scan replacing Aluminium frame with plastic frame on T3 chamber
T7 Chamber used as reference
Slide8MicroMegas chamber with plastic (FR4) frame
FR4 frame
8
MicroMegas Weekly Meeting 2013/03/12
The idea is to monitor the chamber response using different frames.
The metallic frame is close to the R.O. Strips and act like a capacitance and reduce the induced signal.
Alu
frame
Slide9HV scan TH2-T3-T7 Alu
frameMicroMegas Weekly Meeting 2013/03/12
9
As for the HV scan with X-ray gun also in this case a systematic shift of the cluster charge distribution of TH2 w.r.t. T3-T7 is clear evident
500 V
510 V
520 V
530 V
540 V
550 V
Slide10HV scan TH2-T3-T7 Alu
frameMicroMegas Weekly Meeting 2013/03/12
10Cluster Charge distributions of T3-T7 are well reproduced by TH2 applying 10 Volt more, as expected from calibration with X-ray
T3-T7 500 V
TH2 510 V
T3-T7 510 V
TH2 520 V
T7 520 V
TH2 530 V
T3-T7 530 V
TH2 540 V
T3-T7 540 V
TH2 550 V
Slide11T3-TH2 HV with Alu or FR4 frame
MicroMegas Weekly Meeting 2013/03/12
11
Any difference in cluster charge response are shown from T3 and TH2 using aluminum or FR4 frame.
Slide12Maximum Probably Value
of Cluster Charge Distributions12
MicroMegas Weekly Meeting 2013/03/12
The M.P.V. from the Cluster Charge distribution is extracted with Landau function for each step and chamber.No difference for T3 and TH2 chambers are found using FR4 or Aluminium frame.T7 chamber as been used as reference and to normalize the result taking in to account also the variation of the environmental conditions.
Slide13Conclusions
New small (10x10 cm) MM chamber (TH2), with different
thickness of insulating layer between copper and resistive strips have been built and tested.Comparison of the chamber gain, under X-ray flux, with standard TXXX chambers has been performed.The operating point of TH2 seems to shifted of 10 Volts w.r.t. the TXXX chambersNew small cosmic telescope, for systematic studies on 10x10 cm chambers, is now available in the gas lab.
HV scan performed with cosmic on TH2 and T3-T7 chambers confirm the shift on operating point found with the X-ray scan ( This kind of shift can be simply explained with few microns of difference of the amplification region )
Cluster charge response seems to be independent from the frame (
alu
or FR4) adopted
to close the small chambers
13
MicroMegas Weekly Meeting 2013/03/12