No25 September2011 Precautions for Magnetic Resonance No 2011 A patient received a burn Degree III on the inside of the thigh after having an MRI scan Contact of the insides of the thighs duri ID: 942272

mri contact medical skin contact mri skin medical position information gantry pharmaceuticals conductive burns cables risk patients wall 2011

Share:

Link:

Embed:

Download Presentation from below link

Download Pdf The PPT/PDF document "Medical Safety Information Pharmaceutica..." is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.


Presentation Transcript

■Medical Safety Information Pharmaceuticals and Medical Devices Agency No. 25 September, 2011 No25 September2011 Precautions for Magnetic Resonance No . 2011 A patient received a burn (Degree I-II) on the inside of the thigh after having an MRI scan. Contact of the insides of the thighs during the MRI scan may contribute to form an electrically conductive loop. Precautions for burns ( Part 1 ) Key points for safe use () Make sure that there is no skin contact in patients’ arms and legs at positioning. Provide patients with clear instructions not to move (change their body position) while being scanned in the MRI machine. Gantry Do not contact with inner wall of gantry Inner wall of gantry MRI machine Risk of burns! No formation of electrically conductive loop due to skin There is a risk of burns at the point of contact current which is caused by skin contact or contact with the inner wall of the gantry during MRI scans. ■Medical Safety Information Pharmaceuticals and Medical Devices Agenc

y No. 25 September, 2011 Appropriate position Inappropriate position (1) Place body in non-contact with inner wall of gantry! Inappropriate position (2)Arms folded Place arms and legs apart to avoid skin contact!! Hand clenched Precautions for burns (Part 2) 2 Do not put cables/cords of radio frequency (RF) coils and electrocardiogram monitors, etc.in contact with the patients’ skin. Examples of RF coils No formation of electrically with cableRF coil RF coil p Photos by Japan Industries Association of There is a risk of burns or failure of if cables of RF coils, etc. form ■Medical Safety Information Pharmaceuticals and Medical Devices Agency No. 25 September, 2011 Place non-conductive pads Appropriate position dry towels, etc.with a risk of contact. Non-conductive Skin should not contact with cables. Loops should not be formed with cables. Fixed band pad so that they pass for patients who have difficulty in maintaining appropriate position. Please contact marketing authorization holders o

f MRI machines about use of non-conductive pads or fixed bands. AboutthisinformationPMDAMedicalSafetyInformationisissuedbythePharmaceuticalsandMedicalDevicesAgencyforthepurposeofprovidinghealthcareproviderswithclearerinformationfromtheperspectiveofpromotingthesafeuseofpharmaceuticalsandmedicaldevices.Theinformationpresentedherehasbeencompiled,withtheassistanceofexpertadvice,fromcasescollectedasMedicalAccident Information Reports the Japan Council Quality Health Care collected Adverse Drug Information Reports the Japan Council Quality Health Care collected Adverse Drug ReactionandMalfunctionReportsinaccordancewiththePharmaceuticalAffairsLaw.Wehavetriedtoensuretheaccuracyofthisinformationatthetimeofitscompilationbutdonotguaranteeitsaccuracyinthefuture.Thisinformationisnotintendedtoimposeconstraintsonthediscretionofhealthcareprofessionalsortoimposeobligationsandresponsibilityonthem,butisprovidedasasupporttopromotethesafe pharmaceuticals healthcare professionals pharmaceuticals healthcare professiona