Gradients Slice selection Frequency encoding Phase encoding Sampling Data collection Introduction Encoding means the location of the MR signal and positioning it on the correct place in the image ID: 476267
Download Presentation The PPT/PDF document "Encoding and Image Formation" is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.
Slide1
Encoding and Image Formation
Gradients
Slice selection
Frequency encoding
Phase encoding
Sampling
Data collectionSlide2
Introduction
Encoding means the location of the MR signal and positioning it on the correct place in the image
RF at precessional frequency of hydrogen applied at 90
0
to B
0
resonates and flips the NMV into transverse plane.
The individual magnetic moments of hydrogen is put into phase.
The coherent transverse magnetization precesses at Larmor frequency in the transverse plane.Slide3
A voltage (signal) is induced in the receiver coil placed in the transverse plane
This signal has a frequency equal to Larmor frequency of hydrogen (at 1.5 T : 63.86 MHz)
The system must be able to locate the signal spatially in three dimensions, so that it can position each signal at the correct point on the image.
First it locates a slice
.
Then it is
located or
encoded
along both axes of the image.
This task is performed by magnetic gradientsSlide4
Magnetic Gradients
Gradients are alterations to the main magnetic field and are generated by coils of wire located within the bore of the magnet.
The passage of current through a gradient coil induces a gradient magnetic field.
The gradient field either adds to or subtracts from B
0
.
B
0
is altered in a linear fashion.Slide5
Magnetic field strength and therefore the precessional frequency of the nuclei situated in the long axis is deferent and is predictable.
This is called
spatial encoding
A
B
C
negative
positive
9998 G
42.5614 MHz
10000 G
42.57 MHz
10002 G
42.5785 MHz
gradient 1 G per cm
2 cm
2 cmSlide6
X,Y,Z Gradient coilsSlide7
There are three gradient coils (X,Y,Z) situated within the bore of the magnet
Z gradient alters the magnetic field strength along the Z- (long) axis
Y gradient alters the magnetic field strength along the Y- (vertical) axis of the magnet
X gradient alters the magnetic field strength along the X- (horizontal /transverse) axis of the magnet
Y
Z
XSlide8
The
magnetic isocentre
is the centre point of the axis of all three gradients, and the bore of the magnet.
The field strength remains
unaltered
at the
isocentre
isocentre
Z
Y
XSlide9
When a gradient coil is switched on, the magnetic field strength is either subtracted from or added to B
0
relative to the isocentre
The slope of the resulting magnetic field is the amplitude of the magnetic field gradient and it determines the rate of change of the magnetic field strength along the gradient axis.
Steep gradient slopes alter the magnetic field strength between two points more than shallow gradient slopes.
Steep gradient slopes therefore alter the precessional frequency of nuclei between two points, more than shallow gradients slopes
Steep & shallow gradientsSlide10
Slice selection
This is done by
first switching the appropriate gradient coil to alter the field strength and the precessional frequency at points along the corresponding axis, and
then by transmitting a selected band of RF frequencies to selectively excite the nuclei which precess in that particular frequencies.
Resonance of nuclei within the slice occurs because RF appropriate to that position is transmitted
Nuclei situated in other slices does not resonate because their precessional frequency is different.Slide11
Z
-gradient selects axial slices
Y
gradient selects coronal slices
X
gradient selects sagittal slices
Y
Z
XSlide12
Slice thickness
To give each slice a thickness, a
band
of nuclei must be excited by the excitation pulse
The slope of the slice-select gradient determines the difference in precessional frequency between two points on the gradient.
Once a certain gradient slope is applied, the RF pulse transmitted to excite the slice, must contain a range of frequencies to match the difference in precessional frequency between two points
This frequency range is called the bandwidth.
As the RF is being transmitted at this point it is called the
transmit bandwidth
.Slide13
To achieve thin slices, a steep slice select slope and/or narrow bandwidth is applied
To achieve thick slices, a shallow slice select slope and/or broad transmit bandwidth is applied.
Transmit bandwidth
Steep gradient
Shallow gradient
Thin slice
Thick slice
broad Bandwidth
Narrow Bandwidth
Thin slice
Thick slice
slice select gradientSlide14
Gradient strength & slice thickness
Shallow (weaker gradient)
Steeper ( strong) gradientSlide15
The system automatically applies the appropriate
gradient slope
and
transmit
bandwidth
according to the thickness of slice required.
The slice is excited by transmitting RF at the centre frequency corresponding to the precessional frequency of nuclei in the middle of the slice,
The bandwidth and gradient slope determine the range of nuclei that resonate on either side of the centre.
In PracticeSlide16
The gap between the slices is determined by the gradient slope and by the thickness of the slice.
In spin echo pulse sequences, the slice select gradient is switched on during the application of the 90
0
excitation pulse and during the 180
0
rephasing pulse, to excite and rephase
each slice selectively
.
In gradient echo, the slice select gradient is switched on during the excitation pulse only.
90
0
180
0
90
0
Slice select gradientSlide17
Frequency encoding
Once a slice has been selected, the signal coming from it must be spatially located (encoded) along both axes of the image
Locating the signal along the long axis of anatomy is done by a process called
frequency encoding
A gradient
is applied along the
selected axis
The precessional frequency of signal along the axis is therefore altered in a linear fashion.
The signal can now be located along the axis of the gradient according to its frequencySlide18
For frequency encoding of
Coronal & sagittal images – use z gradient
Axial images – use X gradient
Axial images of Head – use Y gradient
A
B
C
Nuclei in column A precess at frequency A
Nuclei in column B precess at frequency B
Nuclei in column C precess at frequency CSlide19
In practice
The frequency encoding gradient is switched on when the signal is received and is often called the
readout gradient
FID
Echo
FID
90
0
90
0
180
0
Frequency encoding gradient
rephasing
dephasing
peak
The steepness of the slope of the frequency encoding gradient determines the size of the anatomy covered ;
Field Of View
(FOV) along the axis during scan.Slide20
Phase encoding
The location of the signal along the remaining third axis is achieved by a process called phase encoding.
This is achieved by applying a gradient along this remaining axis
A gradient is switched on it
alters
the
speed of precession
as well as the
accumulated phase of the nuclei
along their precessional path.
It produces a phase difference or shift between nuclei
positioned along the axis.Slide21
nuclei travel slower
Nuclei travel faster
Loose phase
gain phase
14998 G 63.852 MHz
15000 G 63.86 MHz
15002 G 63.868 MHz
Gradient & phase difference Slide22
When the phase encoding gradient is switched off, the magnetic field strength experienced by the nuclei returns to B
0
and the precessional frequency of all the nuclei returns to the larmor frequency.
However the phase difference between nuclei remains
The nuclei travel at the same speed around their precessional paths, but their phases or positions are different.
This difference in phase between the nuclei is used to determine their position along the phase encoding gradient (axis). Slide23
In practice
The phase encoding gradient is switched on just before the application of the 180
0
rephasing pulse in spin echo sequences.
Phase encoding gradient
90
0
90
0
180
0Slide24
Summary of phase encoding
The phase encoding gradient alters the phase along the
short axis
of the anatomy
In
Coronal
images –
x
gradient
In sagittal images - Y gradientIn
axial images - Y gradientAxial images of brain –
x gradientSlide25
Summary spatial encoding
The slice-select gradient is switched on
during the 90 and 180 pulses in spin echo pulse sequences , and
during the excitation pulse only in gradient echo pulse sequences
The slope of the slice-select gradient determines the slice thickness and slice gap (along with transmit bandwidth)Slide26
The phase encoding gradient is switched on
just before the 180 pulse in spin echo, and
between excitation and the signal collection in gradient echo.
The slope of the phase encoding gradient determines the degree of phase shift along the phase encoding axis.
The frequency encoding gradient is switched on during the collection of the signal
The
amplitude
of the
frequency encoding
gradient and the phase encoding
gradient determines the two dimensions of the FOVSlide27
Gradient timing in spin echo
90
0
90
0
180
0
echo
slice select
slice select
Phase encode
Frequency encode
TRSlide28
Sampling
The signal is collected during the frequency encoding gradient (
readout gradient
)
The duration of readout gradient is called
sampling time
The system
samples up to 1024 frequencies
during sampling time
The rate at which the samples are taken is called the sampling rateSlide29
The
number of samples taken
determines the
number of frequencies
sampled
The range of frequencies is called the
receive bandwidth
f1
f2
f4
f3
f5
f6
Receive bandwidth
Frequency columns in FOV
Frequencies sampled are mapped across the FOV along the frequency axisSlide30
Sampling time, sampling rate and receive bandwidth are linked by a mathematical principle called the
Nyquist theorem
.
It states that any signal must be sampled at least twice per cycle in order to represent or reproduced it acurately.
In addition enough cycles must occur during the sampling time to achieve enough frequency samples ( if 256 samples are to be taken 128 cycles must occur during the sampling time)
Number of cycles occurring per second is determined by the receive bandwidth
Receive bandwidth is proportional to the Sampling rateSlide31
Sampling time is inversely proportional to:
The sampling rate
The receive bandwidth
The receive bandwidth affect the minimum TE ( because the sampling time is changed)
Reducing the receive bandwidth increase the TE (sampling time increases) & vise versa
Usually the receive bandwidth & sampling time are fixedSlide32
Nyquist theorum
Sampling once
Reproduced as a straight line
Sampling twice
Reproduced more accuratelySlide33
Bandwidth versus sampling time
Bandwidth
Sampling time (8 ms)
16,000 Hz
8,000 Hz
128 cycles occur
(256 samples can be taken)
64 cycles occur (only 128 samples can be taken)
If bandwidth is reduced, the sampling time must be increased so that the same number of samples can be takenSlide34
Data collection
Location of individual signals within the image by measuring the number of times the magnetic moments cross the receiver coil (
frequency
), and their position around their precessional path (
phase
)
Frequency shift
Phase shift
1 cycle/s
2 cycles/s
3 cycles/sSlide35
K space
The data information is stored in the computer memory location called the K space. Maximum number of lines are 1024
phase
frequency
+ve
-ve
central
outer
One line is filled for one phase encoding gradientSlide36
Data collection – step 1
During each TR the signal from each slice is phase encoded and frequency encoded.
A certain value of frequency shift is obtained according to the slope of the frequency encoding gradient, which is determined by the size of the FOV.
As the FOV remains unchanged during the scan, the frequency shift value remains the same.
A certain value of phase shift is also obtained according to the slope of the phase encoding gradient
The slope of the phase encoding gradient will determine which line of K space is filled with the data from that frequency and phase encodingSlide37
Phase shift & pseudo-frequency
The system cannot measure the phase values directly
It can measure frequency
The phase shift values are converted to a sine wave
The frequency of this sine wave is called a pseudo-frequency
Different phase shift gradient produce different sine waves with different pseudo-frequencySlide38
The pseudo frequency curve
Phase shift value
timeSlide39
Phase encoding gradient & pseudo frequency
Steeper gradients results in high pseudo frequencies
Shallow gradients results in low frequenciesSlide40
In order to fill out different lines of K space, the slope of the phase encoding gradient must be altered after each TR
With each phase encoding one line of
K
space
is filled
Different lines in K space are filled after every TR
The phase encoding gradient is altered for every TR
In order to complete the acquisition all the lines of selected K space must have been filled
The number of lines that are filled is determined by the number of different phase encoding slopes that are appliedSlide41
K space
Line 1 phase encode 1 frequency/phase data
Line 2 phase encode 2
Line 128 phase encode 128Slide42
Fast Fourier Transform (FFT)
The data in K space is converted into an image mathematically by Fourier Transform.
The receive signal is a composite of multiple signals with different frequencies and amplitudes
The
signal intensity/time domain
is converted to a
signal intensity/frequency
domain
RF intensity
Time
Amplitude
Frequency
Time domain
Frequency domainSlide43
Matrix & FOV
The FOV relates to the amount of anatomy covered
It can be square or rectangular
Image consists of a matrix of pixels
Te number of pixels depends on the number of frequency samples and phase encodings
Matrix = frequency samples x phase encodingsSlide44
Matrix
Coarse matrix 4x4
Fine matrix 8 x 8
4 frequency samples
8 phase samples
8 frequency samples
4 phase samplesSlide45
Data collection - step 2, NSA (NEX)
When all the lines of K space is filled the acquisition is over
But the signal can be sampled more than once with the same slope of phase encoding gradient.
Doing so each line of K space is filled more than once
The number of times each signal is sampled with the same slope of phase encoding gradient is usually called the number of signal averages (NSA) or the number of excitations (NEX).
The higher the NEX, the more data is stored and the amplitude of the signal at each frequency and phase shift is greaterSlide46
Scan timing
Every TR, each slice is selected, phase encoded and frequency encoded.
The maximum number of slices that can be selected and encoded depends on the length of the TR.
E.g.
TR of 500ms may allow 12 slices.
TR of 2000 ms may allow 18 slicesSlide47
TR & number of slices
TR
Slice 1
Slice 2
Slice 3
Slice 4
Slice 1 second TR
90
180
echo
TE
Phase encode 1
Phase encode 2Slide48
The phase encoding gradient slope is altered every TR and is applied to each selected slice in order to phase encode it.
At each phase encode a different line of k space is filled. The number of phase encoding steps therefore affects the length of the scan
E.g. 256 phase encodings require 256 x TR to complete the scan.
The scan time is also affected by the number of times the signal is phase encoded with the same phase encoding gradient slope, or NEX . So,
Scan time = TR x Number of phase encodings x NEXSlide49
K space filling
The negative half of the k space is a mirror image of the positive half.
The polarity of the phase gradient determines whether the positive or negative half is filled
Gradient polarity depends on the direction of the current through the gradient coil
The central lines are filled with data produced after the application of shallow phase encoding gradients
The outer lines are filled with data produced with steep phase encoding gradientsSlide50
The steepness of the slope of the phase encoding gradient depends on the current driven through he coil.
The central lines of K space are usually filled first. (if 256 phase encodings are performed 128 positive lines and 128 negative lines are filled.
The lines are usually filled sequentially either from top to bottom or from bottom to topSlide51
Signal amplitude & phase shift gradient
The shallow phase encoding gradients have smaller phase shifts. The resultant signal therefore has a large amplitude
The steeper phase encoding gradients have larger phase shift along their axis and therefore small signal amplitudesSlide52
Phase encoding slope & signal amplitude
Steeper gradient
medium gradient
shallow gradient
Low amplitude
medium amplitude
high amplitudeSlide53
Signal amplitude & frequency gradient
The vertical axis of k space correspond to the frequency encoding
The left of the k space is a mirror image of the right
The centre represents the maximum signal amplitude because all the magnetic moments are in phase
The magnetic moments on either side are either rephasing and dephasing and therefore the amplitude is lessSlide54
Signal amplitude & frequency gradient
Rephasing
Dephasing
PeakSlide55
K space filling & spatial resolution
Number of phase encodings determines the number of pixels in the FOV along the phase encoding direction
If the FOV is fixed voxels of smaller dimensions result in an image with high spatial resolution
The
steeper gradients
result in
high spatial
resolution (two adjacent points have different phase values and can be differentiated)Slide56
The outer lines of K space contain data with high spatial resolution
The central lines of k space contain data with a low spatial resolution
The
central portion
of k space
contains data
that has
high signal amplitude
& low spatial resolutionThe outer portion of k space
contains data that has low signal amplitude and high spatial resolutionSlide57
Resolution & Amplitude
High spatial resolution
High signal
High spatial resolutionSlide58
Way of filling K space
The
amplitude of frequency encoding
gradient determines
how far to the left and right
K space is traversed and this in turn determines the size of the FOV in the frequency direction of the image
The
amplitude of the phase encoding
gradient determines
how far up and down a line of K space is filled and in turn determines the size of the FOV in the phase direction of the image (or the spatial resolution when the FOV is square)
The polarity of each gradient defines the
direction traveled through K spaceSlide59
K space filling in gradient echo
The frequency encoding gradient switches negatively to forcibly dephase the FID and then positively to rephase and produce a gradient echo
Frequency encoding gradient is negative, k space traversed from left to right
Frequency encoding gradient is positive, k space traversed from right to left
Phase encode gradient is positive , fills top half of K space
Phase encode gradient is negative, fills bottom half of K spaceSlide60
K space filling in gradient echo
Phase encode amplitude determines distance B
B
A
Negative gradient traverse from centre through distance A
Positive gradient traverse from centre through distance C
C
Line of k space filledSlide61
Manipulation of K space filling
The way in which K space is filled depends on how the data is acquired and can be manipulated to suit the circumstances of the scan; e.g. in the following
Rectangular field of view
Anti-aliasing
Ultra fast pulse sequences
Respiratory compensation
Echo planar imagingSlide62
Partial or fractional echo imaging
This refers to when only part of the signal is read (sampled) during application of frequency encoding gradient
As the
sampling time is reduced
minimum
TE can be reduced
This allows maximum T1 and proton density weighting and number of slices for a given TRSlide63
Readout gradient
Minimum TE
Minimum TE reduced
Only half of the k Space is filled
This extrapolated from filled segment
Partial echo imaging
Only this half is readSlide64
Partial or fractional averaging
The negative and positive halves of K space on each side of the phase axis are symmetrical and mirror image of each other
The filling of at least half of the lines is adequate to produce an image
If 60% of lines are to be filled only 60% of phase encodings are required and the remaining lines are filled with zeros
The scan time is there fore reduced
E.g. 256 phase encodings and, 1 TR and ¾ NEX is selected
This is called partial or fractional averaging Slide65
Partial averaging
75% of k space is filled with data
25% is filled with zeros
If phase encodings = 256
TR = 1s
NEX=3/4,
Scan time = 256 x ¾ x 1 = 192 s
If phase encodings = 256
TR = 1s
NEX=1,
Scan time = 256 x 1 x 1 = 256 sSlide66
PRE-SCAN
This is a method of calibration that should be performed before every data acquisition. It includes;
Finding the centre frequency on which to transmit RF. I.e. Resonant frequency of water protons within the area under examination
Finding the exact magnitude of RF that must be transmitted to generate maximum signal in the coil. (to flip the NMV through 90
0
)
Adjustment of the magnitude of the received signal so that it is not too large nor too small.Slide67
Reasons for failing pre-scan
The coil is not plugged in properly
The coil is faulty
Chemical saturation techniques are utilized and there is an uneven distribution of fat and water in the area to be saturated
The patient is either very large or very smallSlide68
Types of acquisition
Sequential
:– data collected for slice by slice (k- space for each slice is filled one by one)
Two-dimensional volumetric
:– data collected for all the slices simultaneously (line 1 in first slice, then line 1 in slice 2)
Three-dimensional volumetric
(
volume imaging
):-collect data from total volume. The excitation pulse is not slice selective, and the whole prescribed volume is excited. At the end of acquisition the volume is divided into partitions by slice select gradient which separates the slices according to their phase value along the gradient. (This is called slice encoding)Slide69
End