Patients should be recommended to start drinking at least 2 L of water a day regularly for hydration Before treatment ask the patient to undress completely and remove all of their belongings with metal ID: 605738

treatment applicator area power applicator treatment power area patient time reflective heat adjust stop sensation multi air screws turn

Share:

Link:

Embed:

Download Presentation from below link

Download Presentation The PPT/PDF document "Patient Preparation" is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.


Presentation Transcript

Slide1

Patient Preparation

Patients should be recommended to start drinking at least 2 L of water a day regularly for hydration.Before treatment, ask the patient to undress completely and remove all of their belongings with metal, e.g., any accessories, watch, belt, mobile phone and so on.Exposure the treatment area.[Note] There could be a possibility of having a panniculitis for certain patient whose nature of fat cell is more easily and quickly to be dissolved in process of apoptosis, so this patient will be treated with lower energy setting than normal parameter for the following remaining treatments.How to adjust the distance of the side cablesArrange the two fixing guides to maintain the proper distance between the two side cables. Avoid to contact the air hose to both side cables.

How to adjust the AB applicator

The applicator should be located directly over the treatment area and placement is recommended to be as close as possible to the target tissue, and at a uniform distance from it. [Tip] Roll up a towel and put it under the patient’s lower spine so that theupper and lower part of the abdomen are as even as possible.Ask the patient to breath naturally to make sure that the applicator does not come in contact with the abdominal skin due to respiration-related abdominal movement. It is important to avoid contact between the applicator and the abdomen to avoid any possibility of burn formation.

1

enCurve Parameters- AB Applicator

Endpoint :

Heat sensation over the entire body during or immediately after the treatment indicates its temperature of deep heat has been increased . Mild erythema can be seen over the treatment area (especially on the flanks)

Before Treatment

Contraindications:

Implanted electronic devices such as cardiac pacemaker, bladder stimulator, spinal cord stimulator. Implants, areas where implants have been removed, damaged implants, and metal inclusions near the treatment area. Thermohyperesthesia, thermohypoesthesia. Pregnant women or possibilities to be pregnant. ※ Please see Operator’s manual for a patient who needs a consultation.

Indication

Power (W)

Time (min.)

Reflective power (W)

Placement of applicator

Treatment interval

No. of sessions

Downsize

Abdominal Circumference

300

20~40

Less than 5

As

close as possible

without touching.

Le

ss than 5 cm

1 week

4~6 Slide2

2

enCurve Parameters- MULTI ApplicatorEndpoint : Heat sensation and mild erythema over the entire treatment area (especially on the flanks).

How to adjust MULTI applicator on the abdomen

The adjustment of Multi applicator on the abdomen is same as what you do with AB applicator.[Note] MULTI applicator can be used to fully cover smaller abdominal area, but the heat sensation might be stronger than AB applicator.How to adjust MULTI applicator on the legsPlace any supportable objects(e.g., roll up towel) under the treatment area, then adjust and wrap the applicator over the treatment area. Make sure the applicator does not come in contact with the surface of the skin.How to adjust MULTI applicator on the armsLet the patient to be sit during the treatment and place supportable objects to put patient’s lower arm on such as a table or a bed.

Make sure the applicator does not come in contact with the surface of the skin.[Caution] Do not directly look at the applicator during the treatment, especially on arms.Treatment position examples with MULTI Applicator

Before Treatment

Indication

Power (W)Time (min.)

Reflective power (W)Treatment intervalNo. of sessionsDownsizeAbdominal Circumference 250~30020 Less than 51 week

4~6 DownsizeLegsCircumference

200~

20 Less than 51 week4~6 DownsizeArmsCircumference250~20 Less than 51 week4~6

Endpoint :

Heat sensation and mild

erythema

over the entire treatment area (especially on the lateral area). Slide3

GUI Monitoring

Impedance synchronization is started by pressing the start button. The energy will be emitted after the sound of a beep for 10 seconds.Check the Current Power and the Reflective Power for 5 minutes at the beginning of the treatment to see if the energy is emitted stably. If frequent variation in reflective power is seen, relocate the applicator or ask the patient to be relaxed.It is important to maintain a reflective power to less than 5 W to maximize delivering deep heat to the fat layer.If the reflective power exceeds over 30 W for more than 4 seconds, the system will be automatically paused to allow impedance matching.Patient MonitoringOnce the device is in operation, the patient will feel a warm sensation on the treatment area which will spread out over the entire body.Some patients may feel discomfort with more heat sensation on a particular area, especially on the lateral area. If the patient can not endure the heat, adjust the distance between the applicator and the skin.

It is recommended to check the reflective power and the heat sensation of the patient every 5 to 10 minutes to maintain a stable and effective treatment procedure.

How to use the patient stop switchIt can be used by the patients whenever they want to stop the operation by themselves.Pull the patient stop switch for 3 seconds. Treatment will automatically be stopped.3During TreatmentPriority of 2 steps to make Reflective Power less than 5W. 1st step: Relocate the applicator to achieve an appropriate distance. 2

nd step: Gradually decrease the power level by 10 W.[Patient stop switch]Slide4

GUI (Graphic

User Interface)4

Power (W)

Energy intensity10 to 300 is available Time (Min)Treatment time1 to 20 minutes setting is availableCurrent Power (W)The amount of power on the treatment area in real-timeRemaining Time (Min)The remaining time for treatment based on the time set in real-timeReflective power (W)The amount of reflected energy from treatment areaForte mode (Optional)Delivering the optimized preset parameter Air Mode (Optional)Air will be blown onto the treatment area when Air mode is activated. Mode 1/2/3 can be selectedPause

OperatingPower buttonDevice On/Off

Emergency Stop buttonStart/Pause/Stop ButtonPause by pressing shortlyStop by pressing for more than 3 sec.

Post Treatment

Treatment in Combination

Applying either HEALITE 633 nm or 830 nm LED-LLLT, or both of them, is recommended to enhance body circulation, and for the activation and production of phagocytic cells. (

e.g., macrophages)Instructions for the Patient

Drink at least 2 L of water daily after the operation including the following day to activate lymphatic drainage of dead fat cells.Mild aerobic exercise will be helpful to improve the clinical outcome.

Mild erythema can be seen for 24 hours.The clinical effect could be seen 2 weeks after the first treatment.Applicator SelectionManually select the applicator alternatively Slide5

5

How to replace the applicatorHow to remove the applicator Turn off the device.Bend the arm of the device in 90 degree.Disconnect the cable from the applicator(Disconnect Air hose for AB Applicator).Turn screws in counter-clockwise. [Caution] Do not pull out screws from the head.Turn the applicator in clockwise to unplug the applicator.

How to connect the applicator

Make its direction of both head of the arms and screws horizontally.Place the applicator in direction of 8 o’clock, and insert the applicator and turn in counter-clockwise until it cannot be turned any more to match the screws with the applicator.Press and turn screws in clockwise. Connect the cables to the applicator(Needed to connect Air hose with AB Applicator)