Dr A A KHINE DIVISION OF CHEMICAL PATHOLOGY NHLS TYGERBERG STELLENBOSCH UNIVERSITY Laboratory Management workshop 36 June 2019 Good laboratory practice requires E rror of measurement must be small enough that a result reflects a patients biological condition ID: 909753
Download Presentation The PPT/PDF document "STANDARDISATION, HARMONISATION AND TRACE..." is the property of its rightful owner. Permission is granted to download and print the materials on this web site for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.
Slide1
STANDARDISATION, HARMONISATION AND TRACEABILITY
Dr A
A
KHINE
DIVISION OF CHEMICAL PATHOLOGY
NHLS TYGERBERG
STELLENBOSCH UNIVERSITY
Laboratory Management workshop 3-6 June 2019
Slide2Good laboratory practice requires
E
rror of measurement must be small enough that a result reflects a patient’s biological condition.
Comparable results that are independent of:
Where and when a test was performed
The measurement procedure used
Slide3Total Error
Calibration bias to an accepted reference ( and also includes bias from calibrator lot-to-lot variability)
Imprecision in a measurement procedure
Sample specific influences
non-specificity for the
measurand
interfering substances
Slide4Why We Need Comparable Results?
If different measurements give different results for the same patient sample:
Clinical practice guidelines become less useful.
Patients may receive incorrect treatment.
Patients may be moving around, thus specimens sent to various labs
Slide5Three areas of standardization in the clinicallaboratory
procedures that are followed before analysis
How sample is analysed (test methods used)
What calibrator is used and how it is traceable
Slide6Standardisation versus harmonisation
Standardization:
Same method or same instrument used
Using calibrator with a traceability ref number
Results are uniform among measurement procedures
Reporting units are the same
Harmonization:
Different methods are used but results are not uniform but comparable
Reporting units may not be the same originally but conversions help to make comparable (SI units)
No reference measurement procedure and no “pure substance” reference material
Slide7How To Achieve Comparable Results?
Laboratories using a common traceable calibrator or different calibrators with the same traceability ref number
Performance is monitored and maintained by surveillance using proficiency testing (PT), external quality assessment (EQA), or a certification program.
Slide8Metrological traceability
True value is not known unless it is attached to a reference so that value can be assigned
is defined as “the property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty”.
Traceability is established to SI using a reference calibrator and reference measurement procedure
Slide9Metrology traceability chain
Slide10Slide11Slide12Slide13COMMUTABILITY ISSUES
Commutable means that
the ref material reacts with the reagents in the ref method and other methods in the same way that analyte in the patient sample would
Thus, values obtained from the
calibration material and
from
native
clinical samples
have the same relationship
across measurement
procedures for
the same
measurand
.
Ideally a reference material should be validated for commutability against all field methods and types of samples but it is rarely the case
Poor commutability means there is matrix effect affecting reaction between reagent and
measurand
Slide14When there is no or poor commutability
A manufacturer uses a “master lot of calibrator” that may be traceable to a non-commutable reference material, then the Traceability chain is broken
Even though manufacturers claim traceability, the process fails to provide equivalent results for patient samples when different measurement procedures are used.
Slide15Slide16Calibration Traceability Does Not Ensure Accuracy for an Individual Patient Sample
Measurement procedure may not be specific for the
measurand
.
Measurand
may not be well defined.
Molecular form(s) of clinical interest
Interfering substances present in a patient’s sample may influence the result
Slide17Use of a Non-commutable Material forCalibration Will Cause
Incorrect value assignment for the manufacturer’s calibrator
Incorrect results for patient samples
Slide18Slide19Slide20Slide21Calibrators in the routine laboratory
A manufacturer's calibrator is proprietary (patented)- intended for use with a specific measurement procedure.
It cannot be used with a different manufacturer’s measurement procedure
This is because the company has validated its commutability only with their own reagent
Slide22Change in Practice Needed
In the business of producing and marketing PRM, SRM and TRM, there must be requirement to submit commutability validation against
Reference methods
Manufacturer’s methods and reagents
Routine clinical laboratory procedures and reagents
A guideline for validating commutability is available: CLSI C53-A-Characterization and Qualification of Commutable Reference Materials for Laboratory Medicine (2010)
Slide23No traceability
No reference measurement procedure or
No reference material
Can there be harmonisation?
Slide24Barriers to Harmonization
Lack of a systematic process to identify and prioritize measurands
Materials labelled as “reference materials” that have not been validated to be commutable for the intended measurement procedures
Inadequate definition of the
measurand
Inadequate analytical specificity for the
measurand
Lack of systematic procedures to implement harmonization, in particular:
When there is no reference measurement procedure
When there is no reference material
Slide25Initiatives in establishing S and H
IFCC has been a key leader in developing science in developing PRMs and RMPs.
Many national and international groups are engaged
JCTLM (joint committee for traceability in lab measurements) composed of three organizations:
International Committee of Weights and Measures
International Federation of Clinical Chemistry and Laboratory Medicine
International Laboratory Accreditation Cooperation
Slide26Functions of JCTLM
Maintain the list of reviewed and approved
Reference measurement procedures
Reference laboratories
Reference materials
Commutability requirements are being addressed.
Slide27Possibilities for consideration:
T
raceable to an all methods mean (outliers removed) of a panel of patient samples
Traceable to a designated measurement procedure (arbitrary, but which has good correlation with clinical outcome)
Slide28References
Mary Lou
Gantzer
, PhD, FACB. President, CLSI, CEO,
BioCore
Diagnostics. Laboratory Standardization: Bringing Order to Chaos
Standardization –the theory and practice-Jill Tate, M
Panteghini
The JCTML- A global approach to promote standardization of clinical laboratory test results
Traceability, Reference systems and result comparability- Clin
Biochem
reviews- 28Aug 2007
ISO 17511
Slide29EXERCISE
What are the analytes in routine chemistry that have been standardized using both ref material and ref method procedure?
What are the analytes that have been standardized through development of primary ref material but not all routine laboratories use the reference method procedure?
What are the analytes that have been standardized through development of primary ref material but lacking method standardization?
What are the analytes that have not been standardized through both development of primary ref material or method standardization?
How has NHLS tried to harmonise the analytes under 2, 3, 4 and 5?