LOT VERIFICATION Labs for Life Project QUALITY

Published  . 0 views
↓ Download
LOT VERIFICATION Labs for Life Project QUALITY
1 / 1
LOT VERIFICATION Labs for Life Project QUALITY - slide 1 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 2 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 3 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 4 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 5 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 6 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 7 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 8 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 9 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 10 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 11 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 12 of 13 LOT VERIFICATION Labs for Life Project QUALITY - slide 13 of 13
Description: LOT VERIFICATION Labs for Life Project QUALITY CONTROL TRAINING What Standards say. ISO 15189: 5.3.2.3 (Reagents and consumables ) Acceptance testing Each new formulation of examination kits with changes in reagents or procedure, or a new

Related Topics

Download Presentation

"LOT VERIFICATION Labs for Life Project QUALITY" is the property of its rightful owner. Permission is granted to download and print the materials on this website for personal, non-commercial use only, and to display it on your personal computer provided you do not modify the materials and that you retain all copyright notices contained in the materials. By downloading content from our website, you accept the terms of this agreement.

Presentation Transcript

slide1. LOT VERIFICATION Labs for Life Project QUALITY CONTROL TRAINING<br>
slide2. What Standards say…. ISO 15189: 5.3.2.3 (Reagents and consumables )
Acceptance testing
Each new formulation of examination kits with changes in reagents or procedure, or a new lot or shipment, shall be verified for performance before use in examinations. Consumables that can affect the quality of examinations shall be verified for performance before use in examinations. NABL 112 :
The laboratory shall check each lot of control and reagent against an earlier tested in-use control / reagent lot. The laboratory can follow guidelines mentioned in Annexure III. Records of comparative data shall be maintained. 7/19/2017 QC Training- Labs for Life Project 2<br>
slide3. Transportation Factors which affect the Quality of Reagents 7/19/2017 QC Training- Labs for Life Project 3<br>
slide4. 7/19/2017 4 Transportation Temperature and Storage temperature Temperature control during transport Temperature Control in your store before issue<br>
slide5. NABL 112 Lot verification is needed for all reagents, Quantitative, Semi quantitative and qualitative analysis

In this video we will discuss only quantitative lot verifications
Please check details for other details 7/19/2017 QC Training- Labs for Life Project 5<br>
slide6. Specify acceptance testing for Qualitative and Quantitative tests
For SQCs
Decide the acceptance limits
The MDP of each analyte where verification is needed
The material you will use for Lot Verification
Quality Specifications eg..TE< TEa/ Sigma> 4 Do you have a QSP? 7/19/2017 QC Training- Labs for Life Project 6<br>
slide7. (CLSI EP 26 A) Lot to lot Verification can be done by: Using Sample

Using QCs 7/19/2017 QC Training- Labs for Life Project 7<br>
slide8. 7/19/2017 QC Training- Labs for Life Project 8 TE/ Sigma<br>
slide9. Samples to be done in triplicate with the old and new lots (CLSI EP 26 A)
Understand the MDPs
Understand acceptable percent differences If You are using sample for lot verification 7/19/2017 QC Training- Labs for Life Project 9<br>
slide10. If You are using QCs for lot Verification Can set better quality specifications like TE and Sigma
However,

Matrix Effect may be an issue especially in tests like immunoassays 7/19/2017 QC Training- Labs for Life Project 10<br>
slide11. Method: Using Samples Run the same sample (individual or pooled) in triplicate using old and new lot reagents
Find the percent difference between each run using the formula
% Difference= (I new lot value- old lot value I)/ Old lot value *100
Set acceptance limits ( eg; 1SD at that level / < 10 %) 7/19/2017 QC Training- Labs for Life Project 11<br>
slide12. 7/19/2017 QC Training- Labs for Life Project 12 Method: Using QC Using the new lot reagent run the QCs, at all levels, 8 times each
Find the mean and SD
Using the 4 key numbers, Mean, Target, SD and TEa calculate TE and Sigma
Target value ( is the Mean from old reagent/calibrator lot)
Calculate Sigma performance
Judge Acceptability as per Quality Specifications<br>
slide13. Thanks 7/19/2017 QC Training- Labs for Life Project 13<br>